ZR-6010 Aerosol Photometer

Short Description:

Aerosol photometer is designed base on Mie scatter principle, which is used to test whether there is leakage on HEPA filter. 


  • Sampling flowrate: 28.3L/min
  • Concentration detection range: (0.0001~125)μg/L
  • Size: (length 380×width 400×height 170)mm
  • Weight: About 8kg
  • Power consumption: <100W
  • Product Detail

    Specification

    Aerosol and Dust Monitors conforms to related national and industry standard, could realize rapid detection of upstream and downstream concentration detection and real-time display leakage on the host and handheld device, and could find leaking position rapidly and accurately. It is applicable to leakage detection of clean room, VLF bench, biosafety cabinet, glove box, HEPA vacuum cleaner, HVAC system, HEPA filter, negative pressure filtering system, operating theatre, nuclear filter system, collection protect filter.

    Standards

    GMP                        Factory and device

    ISO14644-3:2005   Cleanrooms and associated controlled environments—Part 3:Test methods

    GB 50591-2010      Code for construction and acceptance of cleanroom   

    YY0569-2005          Biosafety cabinet

    NSF49-2002           Biosafety cabinet

    Features

    > Long life laser light source;

    > Detected by high precision photomultiplier;

    > Support PAO and DOP multiple aerosol type;

    > Pixel and matrix colored display;

    > Equipped special handheld device, realize control, display and sampling function;

    > Large capacity data storage, real-time save sampling data;

    > Automatic light and voice alarm after exceeding set value;

    > Support export historical data to USB flash disk or print with thermal printer;

    > Real-time print detection data like leakage and so on;

    > Sampling data could be imported to PC via special software;

    > Automatic failure detection and protection.

    Deliver Goods

    deliver goods Italy
  • Previous:
  • Next:

  • The Main Parameters Parameter Range Maximum Permitted Error(MPE)
    Sampling flow 28.3L/min ±5%
    Concentration detection rage (0.0001~125)μg/L
    Leakage detection 0.0001%~100%
    Detection precision 0.01%~100%   resolution is 1%
    Detection repeatability 0.01%~100%  resolution is 0.5%
    Data storage capacity 1000 groups
    Power adapter Input AC100~240V 50/60Hz   Out put DC24V 6.67A
    Host size (Length 380×width 400×height 170)mm
    Overall weight About 8kg
    Overall power consumption <100W

    Conditions for running and storage:

    The Main Parameter Parameter range
    Environment temperature (10~35)℃
    Environment humidity 5%-85%(No condensation、No icing)
    Storage requirements (-10-40)℃ No condensation when relative humidity lower than 85%.